Reach In Plant Growth Chamber - E-36L2

Percival® Scientific specializes in products to make research of all types in all fields easier, faster, more accurate and more intuitive. The E-36L2 Seed Germination Chamber is designed to make plant research easier, including the study of lighting conditions on vascular plant development, how to optimize plant production, studies of plant pathology, seedling germination, plant morphology, phytochemistry and much more. Percival Scientific’s IntellusUltra controller gives you detailed and continuous control over the chamber’s internal environment, with flexible standard controls and customizable additions that give you extra specificity. Take a look at the product specifications to compare with other germination chambers and find the best fit for your needs.

Download a PDF of this product – E-36L2

Description

Applications
-Frequently used for research applications such as lighting for vascular plants to facilitate standard plant production, plant pathology research and seedling germination and development
-Many other applications exist for this product

Percival’s IntellusUltra Controller
-Percival Scientific has built a reputation of providing flexible, customized options for research scientists around the world. We have taken the philosophy to the next level with our improved IntellusUltra Controller. Now choose from the levels of functionality that meet your needs.

Lighting System
-Each tier of shelves is lighted by cool white fluorescent lamps and incandescent lamps properly spaced for uniform light intensity
-Intensity programmable up to 680 micromoles/m2/sec of light irradiance measured at 6” from lamps on 3 on/off light events
-Programming and control of the lighting is done via IntellusUltra real time controller

Airflow/Circulation
-Air circulation inside chamber is from a specifically designed air diffuser (air travels along the entire back wall, over the shelves and returns to the ceiling fans through an opening between the light fixtures and the doors)

Cabinet Construction
-Interior constructed of 22-gauge electro-zinc plated steel
-Exterior constructed of 18-gauge exterior electro-zinc plated steel
-Welded seams and joints on outer and inner shells
-Inner shell supported by non-compressing/non-thermal conducting material locking inner liner in place without a metal-to-metal bond to outer case
-Chamber is completely self-contained

Insulation
-Woodless construction using CFC free insulation (overall wall thickness is 2”, ample insulation for maintenance of stated temperature range)

Door
-One door opening of 29.2” x 57.5” providing full access to chamber interior (magnetic gasket provides a tight seal to door frame)

Interior Space
-29.7 cu.ft. with work area of 10.8 sq.ft. provided on two tiers

Shelving
-Two tiers of white epoxy coated steel wire shelving (each shelf is 27”D x 28.8”W)
-Shelves are supported by shelf clips allowing 1/2” vertical adjustments
-Maximum clearance between shelves is 22.6” per tier with both shelves installed

Finish
-Interior and exterior painted with high reflective, environmentally friendly, high temperature baked white powder coating

Refrigeration
-Self-contained air-cooled condensing unit with hot gas bypass system for continuous compressor operation, extended life and close temperature control (this continuous running condensing unit ensures precise temperature control by alternating cycling refrigerant and hot gas to coil; this also prolongs life of compressor, and eliminates risk of ice buildup in coil)
-Solenoid valves have extended stem for quiet and long life operation
-Ceiling mounted evaporator coil incorporates twin air circulation fans in aluminum housing (heat rejection to ambient with standard chamber: 5900 BTU/hr)

Temperature Range
– 7-44C (+/- 0.5C) lights on and 2-44C (+/- 0.5C) lights off

Temperature Safety Limit Controls
-Experiment Protection: Adjustable high and low temperature controls, audible alarms, and visual indicators provided
-Controls shut down all power to chamber, activating alarms (when the temperature returns to the normal range the system will automatically reset)

Convenience Receptacles
-Two 120/1/60 convenience receptacles provided inside chamber

Electrical Requirements
-Consult Geneva Scientific for electrical requirements and amperage draw.

Options
-Additive Humidity Control with Sensor
-Dehumidifier with Sensor
-IntellusUltra Connect
-Android-Based Touch Screen
-CO2 Enrichment Package
-Self-Contained Water-Cooled Condensing Unit
-Dry Alarm Contacts
-Dimmable Lighting
-LED Lighting in Lieu of Fluorescent Lamps
-Please contact us for more optional items

***Specifications subject to change.***

ModelTemp Range
(with lights on)
Interior Space (Volume)Total Shelving Floor AreaMaximum Growing HeightExterior DimensionsLight Intensity (6" from lamps unless noted)# of Tiers
E-36L27-44±0.5 °C29.7 ft3
0.8 m3
10.8 ft2
1.0 m2
22.6 in
57.5 cm
33.5 in (W)
33.6 in (D)
77.2 in (H)
680 µmol/m2/s2