Barrier isolation system provide inherently superior sterility compared to open front clean air devices such as laminar flow clean benches and Class II biological safety cabinets. USP797 guidelines specify that isolators may be situated in an area subject to less severe environmental controls compared with open front clean air devices.
When used as part of a system that includes operator aseptic technique training, process validation, expiration setting, product quality maintenance after the CSP leaves the pharmacy, isolators are a cost-effective solution especially for lower-volume pharmacies. They reduce operating and renovation costs, take up less space, and are easier to maintain.
The positive pressure HPI model is suitable for work involving non-hazardous materials. The work zone and pass-thru interchange are under positive pressure to the room in order to maintain sterility in case of a breach in the barrier isolation system. Negative pressure models should be selected for antineoplastic or cytotoxic compounding applications.
| Model | GPI-4-PS | GPI-6-PS | |
| Add to Quotes | Add to Quotes | ||
| External Dimensions | Without Base Stand | 52.8" x 32.3" x 51.9" 1340 x 820 x 1320 mm |
76.8" x 32.3" x 51.9" 1950 x 820 x 1320 mm |
| With Base Stand (Min) | 59.9" x 32.3" x 78.5" 1522 x 820 x 1995 mm |
83.9" x 32.3" x 78.5" 2132 x 820 x 1995 mm |
|
| With Base Stand (Max) | 59.9" x 32.3" x 88.6" 1522 x 820 x 2250 mm |
83.9" x 32.3" x 88.6" 2132 x 820 x 2250 mm |
|
| Main Chamber Work Zone | 33.1" x 24.0" x 26.4" 840 x 610 x 670 mm |
57.1" x 24.0" x 26.4" 1450 x 610 x 670 mm |
|
| Pass Through | 13.9" x 24.0" x 26.4" 355 x 610 x 670 mm |
13.9" x 24.0" x 26.4" 355 x 610 x 670 mm |
|
| Work Zone and Interchange Chamber Performance | ISO Class 3 (Class 1, Federal Standard 209E) | ||
| Inlet Filter Type | 80% efficiency filter | ||
| Downflow and Exhaust Filter Type | ULPA filter with integral metal guards and filter frame gaskets; fully compliant with EN 1822 (H14) and IEST-RP-CC001.3 requirements (each cabinet has individual downflow and exhaust filters) | ||
| Fluorescent Lamp Intensity | > 950 Lux (> 88 foot candles) | ||
| Isolator Construction | Main Body | 1.2 mm (0.05") 18 gauge electrogalvanised steel with white oven-baked epoxy antimicrobial powder-coated finish | |
| Work Tray | 1.5 mm (0.06") 16 gauge stainless steel, type 304, with 4B finish | ||
| Side Walls | 1.2 mm (0.05") 18 gauge stainless steel, type 304, with 4B finish | ||
Hospital Pharmacy Isolator is available as a standard bench top unit. Additional accessories are available for further enhancement.
| Standards Compliances | Cabinet Performance | Air Quality | Filtration | Electrical Safety |
| CETA CAG-001-2005, USA CETA CAG-002-2006, USA USP Chapter 797, USA |
ISO 14644.1, Class 3, Worldwide JIS B9920, Class 3, Japan JIS BS5295, Class 3, Japan US Fed Std 209E, Class 1 USA |
EN-1822, Europe IEST-RP-CC001.3, USA IEST-RP-CC007, USA IEST-RP-CC034.1, USA |
EN 61010-1, Europe IEC 61010-1, Worldwide |