Labconco Cell Logic Class II Type A2 Biosafety Cabinet Model 342491102
4-Foot | Scope-Ready™ with Temp-Zone™, UV Lamp, Service Fixture, Vacu-Pass™ Portals, and Base Stand
The Labconco Cell Logic Class II Type A2 Biological Safety Cabinet (Model 342491102) delivers personnel and product protection for biosafety level 1 through 4 applications requiring microscopy access and temperature-controlled work surfaces—all within a compact 4-foot footprint.
This configuration combines the Scope-Ready™ package with the Temp-Zone™ temperature-controlled work area, plus UV light, service fixtures, Vacu-Pass™ portals, and a factory-supplied base stand, making it ideal for space-constrained labs that still require advanced functionality.
Containment, Filtration & Exhaust Configuration
-
Cabinet Type: Class II, Type A2
-
Filtration: HEPA-filtered supply and exhaust air
-
Exhaust Options:
-
HEPA-filtered air exhausted back into the laboratory
-
Can be canopy-connected to a facility exhaust system
-
When canopy-connected, this cabinet may be used for applications involving minute quantities of volatile toxic chemicals and trace radionuclides, subject to institutional safety protocols.
Airflow Performance
-
Air Inflow Velocity: 105 fpm (0.53 m/sec)
-
Air Downflow Velocity: 55 fpm (0.28 m/sec)
These airflow rates support consistent containment at the sash opening and uniform laminar downflow across the work surface.
Scope-Ready™ Package
Engineered for microscopy-based workflows:
-
Sash Cutout: 7.0″ W × 10.5″ H (18 × 27 cm)
-
Pure-Vu™ Seal with powder-coated steel collar
-
Maintains containment integrity while accommodating microscope optics
Note: A Stand-Still™ Isolation Platform is required for microscope use and is sold separately.
Temp-Zone™ Temperature-Controlled Work Area
Supports temperature-sensitive procedures:
-
Work Area Size: 10.5″ W × 12.0″ D (27 × 30 cm)
-
Location: Left side of the work surface
-
Water Circulator Required (Not Included)
Circulator Requirements:
-
Minimum flow rate: 0.4–0.5 GPM (1.5–1.9 LPM)
-
Typical volume: 6 liters
-
Operating temperature range: 0–50 °C (chiller/heater)
Accessory Package
-
UV Lamp
-
Right-Side Service Fixture (air, vacuum, etc.)
-
Right and Left Vacu-Pass™ Portals for routing cords and tubing while maintaining containment
Dimensions & Physical Specifications
-
Nominal Width: 4 ft
-
Sash Opening Width: 22.5″ (57 cm)
-
Base Stand: Included
Electrical Requirements
-
Voltage: 115V
-
Frequency: 60 Hz
-
Current: 16A
Compatible with standard North American laboratory electrical service.
Typical Applications
-
Biosafety workflows requiring microscopy and temperature control
-
Cell culture and biological research
-
Laboratories needing advanced features in a compact footprint
-
BSL-1 through BSL-4 environments (per institutional approval)
Need Help Choosing a Biological Safety Cabinet?
Selecting the right biosafety cabinet depends on your workflow, biosafety level, lab layout, and certification requirements. Geneva Scientific’s experts can help you compare models, identify accessory options, and ensure the best fit for your needs.
Contact us for a quote